| Safety standard |
IEC 61010-1, IEC 61010-2-032 Installation CategoryⅡ600V or CategoryⅢ300V. |
| E.M.C. standard |
IEC 61326 |
| Measuring method |
Dual slope integration mode |
| Display |
3.5 digit LCD, max. reading of 1999 |
| Jaw opening capability |
55mmφ |
| Over range indication |
Blanking of all digits except MSD 1 |
| Low battery indication |
“Battery” mark on LCD |
| Data hold indication |
“DH” mark on LCD |
| Sampling |
2 times/sec |
| Withstanding voltage |
AC 3700V 1 minute max. (Between the core of CT and outer case) |
| Operating temperature |
0℃ to 40℃, 80%RH max. ( Without condensation ) |
| Storage temperature |
-10℃ to 60℃, 70%RH max. ( Without condensation ) |
| Power supply |
1.5V (AM-4, LR03 or AAA)×2 |
| Power consumption |
3.5mW |
| Auto power off |
Approx. 10 minutes later after power on |
| Battery life |
Approx. 500 hours continuous |
| Size |
85(W)×240(H)×34(D)mm |
| Weight |
Approx. 350g |
| Accessories |
Hard carrying case
Instruction manual
Test lead
Batteries |
| Accuracy |
| Function |
Range |
Accuracy |
Maximum input |
AC A (50/60Hz)
Manual ranging |
20A |
±1.2% rdg ±10 dgt |
AC2000A
(30 second) |
| 200A |
±1.2% rdg ±10dgt |
| 2000A |
±1.2% rdg ± 8 dgt |
AC V (50/60Hz)
DC V
Auto ranging |
2V |
±0.7% rdg ± 5 dgt |
DC 750V
or
AC 750V rms |
| 20V |
±1.2% rdg ± 5 dgt |
| 200V |
±1.2% rdg ± 5 dgt |
| 750V |
±1.2% rdg ± 5 dgt |
Resistance(Ω)
Auto ranging |
200Ω |
±1.2% rdg ± 5 dgt |
Input protection
250V rms
(30 second) |
| 2kΩ |
±1.2% rdg ± 5 dgt |
| 20kΩ |
±1.2% rdg ± 5 dgt |
| 200kΩ |
±1.2% rdg ± 5 dgt |
| 2000kΩ |
±1.2% rdg ± 5 dgt |
| 20MΩ |
±3.0% rdg ±10 dgt |
| Continuity check |
2kΩ |
Beeper <approx.300Ω |
250V rms |
| Diode test |
2V |
±10% rdg ± 3dgt |
250V rms |
|